Introduction
2 - 33
Options
Device Setup
The
Device Setup
tab in the
Options
view allows you to change the settings of
all test cards of the test procedure running on
CPC 100
and all test cards
inserted afterwards. The test cards can be instantiated in different ways and for
different purposes. Table 1: ”Instantiation procedures” below displays the
instantiation procedures and the associated default values.
Table 1:
Instantiation procedures
Because the settings of a test card containing results cannot be changed, the
settings made at the
Device Setup
tab in the
Options
view apply only to new
test cards if the currently loaded procedure contains results. To change the
settings of a test procedure containing results (for example ,after pressing the
K
E E P
R
E S U L T S
menu key or after loading a file), clear the results by pressing
the C
L E A R
R
E S U L T S
menu key in all test cards before setting the test
procedure in the
Options
view.
It is strongly recommended to set the test cards using the
Options
view. In well-
founded cases, you can alternatively set a test card individually using the
S
E T T I N G S
menu key (see ”Settings Page” on page 28 of this chapter).
Note:
Changes made in the
Options
view are automatically saved when the
view is changed (for example when
Options
is left to change to
Test
Card View
).
Alternatively, changes can be saved instantly by pressing S
A V E
O
P T I O N S
.
Upon a power supply loss or an accidental switching off of
CPC 100
,
unsaved changes are lost, and the old values are restored.
Procedure
Default values
New test card
Booster, clamp and default frequency
set in the
Options
view
New test card saved as default
Booster and clamp set in the
Options
view
New test procedure from template
(*.xmt)
Booster, clamp and frequency remain
as defined in the template
New procedure from report (*.xml)
Booster, clamp and frequency remain
as defined in the report
Summary of Contents for CPC 100
Page 12: ...CPC 100 V1 41 x...
Page 28: ...CPC 100 V1 41 1 16...
Page 90: ...CPC 100 V1 41 3 14...
Page 194: ...CPC 100 V1 41 6 30...
Page 250: ...CPC 100 V1 41 8 32 Figure 24 Settings of Amplifier test card for this example use case...
Page 258: ...CPC 100 V1 41 9 4...
Page 264: ...CPC 100 V1 41 10 6...
Page 282: ...CPC 100 V1 41 12 10 Figure 9 Saving tests with the CPC Editor...
Page 284: ...CPC 100 V1 41 12 12...
Page 312: ...CPC 100 V1 41 14 12...
Page 316: ...CPC 100 V1 41 15 4...
Page 350: ...CPC 100 V1 41 16 34...
Page 372: ...OMICRON Contact Addresses 22...