Changing Fuses
• Turn off
CPC 100
, unplug the power cord and/or press the Emergency Stop
button.
• We recommend to wait for about 30 seconds. This time is necessary for the
internal electrolytic capacitors to fully discharge.
• Ground the test object, and disconnect it from
CPC 100
. By disconnecting it
you prevent a possibly faulty test object feeding power back into
CPC 100
.
• Locate the blown fuse on the front panel of
CPC 100
, and replace it:
–
6.3A T
(6.3 Amps slow-acting wire fuse 5x20mm) for
AC OUTPUT
in 6A
operation mode
1
or for
DC OUTPUT
.
–
3.15A T
(3.15 Amps slow-acting wire fuse 5x20mm) for
AC OUTPUT
in
130V operation mode.
–
10A FF
(10 Amps very-quick-acting wire fuse 5x20mm) for measuring
inputs.
1. For detailed information about the difference between the 6A and the 130V operation mode refer
to ”Functional Components of CPC 100” on page 2-2.
Note:
Replace with identical fuse type only.