CPC 100 V1.41
16 - 8
2kV AC output
While using the
2kV AC
output, disconnect any cables that may be plugged in
to any of the high-current outputs (
800A AC
or
400A DC
).
Table 4:
Technical data of
2kV AC
output - all
ranges
Table 5:
Technical data of
2kV AC
output at a
voltage range of 2000V
2kV AC output - all ranges
Connection
High-voltage sockets
Insulation
Insulated against all other inputs and outputs.
Voltage with reference to earth must not exceed
2000V
rms
and 2800V
peak
on either socket.
Reinforced insulation against mains power supply, all
interfaces and other safety low-voltage circuits.
Measurement
Internal precision current and voltage measurement,
integrated in
CPC 100
(for details, see tables below).
Abs. max. rating: limited through output
Connection:
internally connected
2kV AC output at a voltage range of 2000V
Maximum current
Peak current I
peak
< 3.0A
Effective current I
rms
< 1.25A
Output power at:
typical apparent
power S
max
typical time t
max
frequency
2000V / 1.25A
2000V / 1A
2000V / 0.5A
2500VA
2000VA
1000VA
60s
130s
> 2hrs
50…200Hz
50...200Hz
50...400Hz
Summary of Contents for CPC 100
Page 12: ...CPC 100 V1 41 x...
Page 28: ...CPC 100 V1 41 1 16...
Page 90: ...CPC 100 V1 41 3 14...
Page 194: ...CPC 100 V1 41 6 30...
Page 250: ...CPC 100 V1 41 8 32 Figure 24 Settings of Amplifier test card for this example use case...
Page 258: ...CPC 100 V1 41 9 4...
Page 264: ...CPC 100 V1 41 10 6...
Page 282: ...CPC 100 V1 41 12 10 Figure 9 Saving tests with the CPC Editor...
Page 284: ...CPC 100 V1 41 12 12...
Page 312: ...CPC 100 V1 41 14 12...
Page 316: ...CPC 100 V1 41 15 4...
Page 350: ...CPC 100 V1 41 16 34...
Page 372: ...OMICRON Contact Addresses 22...