CPC 100 V1.41
16 - 10
Table 6:
Technical data of
2kV AC
output at a
voltage range of 1000V
2kV AC output at a voltage range of 1000V
Maximum current
Peak current I
peak
< 6.0A
Effective current I
rms
< 2.5A
Output power at:
typical apparent
power S
max
typical time t
max
frequency
1000V / 2.5A
1000V / 2A
1000V / 1A
2500VA
2000VA
1000VA
60s
130s
> 2hrs
50…200Hz
50...200Hz
50...400Hz
Measurement
Voltage
Voltage and current
Range:
0 ... 1000V AC
Accuracy:
typ. error <± 0.05%(rd) ± 0.05%(fs),
guaranteed <± 0.1% (rd) ± 0.1% (fs)
Possible crest factor: < 1.5
Phase tolerance
at full scale:
typ. error < ± 0.15°,
guaranteed < ± 0.3°