Introduction
2 - 7
They provide an output voltage of either
• 0 ... 0.5kV with a maximum current of 5A, or
• 0 ... 1kV with a maximum current of 2.5A, or
• 0 ... 2kV with a maximum current of 1.25A.
For details refer to ”High-Current and High-Voltage Outputs” on page 16-5.
Both the voltage of this output and the current that flows from this output are
measured internally by
CPC 100
, and displayed on the respective test card.
All outputs are overload and short-circuit-proof and protected against external
high-voltage signals as well as over-temperature.
Note:
The length of time a high current or a voltage can be applied may be
limited due to a high temperature occurrence of the internal
transformer windings and/or heat dissipaters. This high temperature
occurrence may even be increased if
CPC 100
is operated in a hot
environment or is exposed to direct sunlight.
It is highly recommended not to exceed the typical duration time t
max
specified for each current and voltage output (refer to ”CPC 100
Outputs” on page 16-5). If you do, and
CPC 100
develops too high an
internal temperature, the outputs are shut off automatically.
Summary of Contents for CPC 100
Page 12: ...CPC 100 V1 41 x...
Page 28: ...CPC 100 V1 41 1 16...
Page 90: ...CPC 100 V1 41 3 14...
Page 194: ...CPC 100 V1 41 6 30...
Page 250: ...CPC 100 V1 41 8 32 Figure 24 Settings of Amplifier test card for this example use case...
Page 258: ...CPC 100 V1 41 9 4...
Page 264: ...CPC 100 V1 41 10 6...
Page 282: ...CPC 100 V1 41 12 10 Figure 9 Saving tests with the CPC Editor...
Page 284: ...CPC 100 V1 41 12 12...
Page 312: ...CPC 100 V1 41 14 12...
Page 316: ...CPC 100 V1 41 15 4...
Page 350: ...CPC 100 V1 41 16 34...
Page 372: ...OMICRON Contact Addresses 22...