CPC 100 V1.41
7 - 12
Measurements
IDC:
actual test current from
6A DC
output
VDC:
voltage that IDC generates at the transformer’s secondary
winding, measured at
V DC
input
R meas:
transformer’s winding resistance, calculated from VDC / IDC
Time:
total elapsed test time
Dev:
deviation in % between the maximum and the minimum
measured values evaluated within the last 10 seconds of the
measurement. The results are considered stable if Dev < 0.1%.
Press the context-dependent K
E E P
R
E S U L T
menu key to
save the actual result in the measurement table. This adds a
new line to the measurement table and the next measurement
can be started.