The "Measure burden" option
Select the check box "Measure Burden" to measure the burden in VA.
Figure 4:
Setup for testing ratio,
polarity and burden with
injection to current input
Note:
This option is only useful as long as the injected current Itest is about
of the magnitude of the nominal current Iprim.
Note:
For CTs that have a hole rather than a primary busbar just loop the
high-current cable through the hole and connect the high-current
clamps together.