CPC 100 V1.41
16 - 32
Environmental Conditions
Climate
Table 19:
Environmental
conditions - climate
Shock and Vibration
Table 20:
Environmental
conditions - shock and
vibration
Climate
Operating temperature
-10° ... +55°C (+14 ... +131F)
Storage and transportation
-20° ... +70°C (-4 ... +158F)
Humidity
5 ... 95% relative humidity, no condensation
Drip-water protection
IP22, tested according to EN 60529/A1:2000
(13.2, 13.3, 14.2.2), not operating, upright
position, that is, front panel must be
horizontal.
IP22 means drip-water protection. Never
operate
CPC 100
with a wet front panel or
wet sockets.
Make sure no water can come in through the
ventilation slots.
Note:
Since the AC and DC output sockets of
CPC 100
can carry life-
hazardous voltage potential and provide life-hazardous currents, we
strongly advise you not to carry out tests with a wet
CPC 100
.
Before switching on
CPC 100
, make sure it is completely dry.
Shock and vibration
Vibration
• Tested according to IEC68-2-6 (operating
mode)
• Frequency range 10 ... 150Hz
• Acceleration 2g continuous (20 m/s²);
10 cycles per axis
Shock
• Tested according to IEC68-2-27 (operating
mode)
• 15g / 11ms, half-sinusoid, each axis
Summary of Contents for CPC 100
Page 12: ...CPC 100 V1 41 x...
Page 28: ...CPC 100 V1 41 1 16...
Page 90: ...CPC 100 V1 41 3 14...
Page 194: ...CPC 100 V1 41 6 30...
Page 250: ...CPC 100 V1 41 8 32 Figure 24 Settings of Amplifier test card for this example use case...
Page 258: ...CPC 100 V1 41 9 4...
Page 264: ...CPC 100 V1 41 10 6...
Page 282: ...CPC 100 V1 41 12 10 Figure 9 Saving tests with the CPC Editor...
Page 284: ...CPC 100 V1 41 12 12...
Page 312: ...CPC 100 V1 41 14 12...
Page 316: ...CPC 100 V1 41 15 4...
Page 350: ...CPC 100 V1 41 16 34...
Page 372: ...OMICRON Contact Addresses 22...