CPC 100 V1.41
16 - 6
Figure 2:
Typical maximum
output voltage of
800A AC
output
depending on frequency
and current
Measurement
Internal precision current measurement from
0 ... 800A AC
Abs. max.
rating:
limited through output
Accuracy:
typ. error < ± 0.1% (rd) ± 0.1% (fs),
guaranteed < ± 0.2% (rd) ± 0.2% (fs)
Possible crest factor: < 1.5
Phase
tolerance at
full scale:
typ. error < ± 0.1°,
guaranteed < ± 0.2°
Resolution:
Measurement
Display
100 mA/bit
10 mA
Connection: internally connected
800A AC output
Note:
The signal for the
800A AC
output is generated by an internal
transformer. To avoid saturation, sudden magnitude changes are
suppressed and the signal is output as a ramp. For several
applications the magnitude change is deferred even more, for
example, for CT ratio measurements.
The ramp changes for a full magnitude are
• one period of the signal, regardless of the size of the change for an
increase ramp
• 300ms for a decreasing ramp, or accordingly less if the magnitude
change is only a fraction of its full value.
voltag
e in V
rms
frequency in Hz
Summary of Contents for CPC 100
Page 12: ...CPC 100 V1 41 x...
Page 28: ...CPC 100 V1 41 1 16...
Page 90: ...CPC 100 V1 41 3 14...
Page 194: ...CPC 100 V1 41 6 30...
Page 250: ...CPC 100 V1 41 8 32 Figure 24 Settings of Amplifier test card for this example use case...
Page 258: ...CPC 100 V1 41 9 4...
Page 264: ...CPC 100 V1 41 10 6...
Page 282: ...CPC 100 V1 41 12 10 Figure 9 Saving tests with the CPC Editor...
Page 284: ...CPC 100 V1 41 12 12...
Page 312: ...CPC 100 V1 41 14 12...
Page 316: ...CPC 100 V1 41 15 4...
Page 350: ...CPC 100 V1 41 16 34...
Page 372: ...OMICRON Contact Addresses 22...