Technical Data
16 - 13
Current Ranges
(automatic range
selection):
0 ... 5A AC
0 ... 500mA AC
0 ... 50mA AC
1
Accuracy:
5A:
typ. error
< ± 0.2% (rd) ± 0.05% (fs),
guaranteed
< ± 0.4% (rd) ± 0.1% (fs)
500mA:typ. error
< ± 0.05% (rd) ± 0.05% (fs),
guaranteed
< ± 0.1% (rd) ± 0.1% (fs)
50mA
1
: typ. error
< ± 0.1% (rd) ± 0.1% (fs),
guaranteed
< ± 0.2% (rd) ± 0.2% (fs)
Possible peak value: ± 20A
Resolution:
Measurement Display
5A:
1 mA/bit
500mA: 100
μ
A/bit
50mA
1
: 10
μ
A/bit
1
μ
A
Phase tolerance
at full scale:
typ. error < ± 0.1°,
guaranteed < ± 0.2°
1. Available for the
CPC 100
serial number Dxxxxx and higher.
2kV AC output at a voltage range of 500V