Chapter 17 Troubleshooting
17-72
Checking the image
quality
Refer to
“5.4 Checking Image Quality”, “5.5 Checking Density Linearity”, and “5.6
Creating QC Pattern Base” in the installation manual
, and check the image quality.
Caution
When replacing the compact flash, always perform “5.5.3 Internal
Densitometer Correction”. However, if there are no densitometers in
the facility, you will need to take it for correction. When aligning with a
densitometer in the facility, use the facility densitometer internally.
Set the logo
Refer to
“5.9 Logo Setup” in the installation manual
to set the logo to be output on a film.
Backup the setup
Refer to
“Chapter 6: Backing up Setup” in the installation manual
to backup the setup
data using the MainteTool.
Creating duplicate data
Refer to
“15.17 Product Setup (Page 15-37)”
and note the following numerical values.
•
SERIAL NO.
•
SIZE FINE TUNING (H)
•
SIZE FINE TUNING (V)
•
SIZE FINE TUNING (Requested size)
END