Chapter 17 Troubleshooting
17-90
Replacing the DRYPRO 832 (V1.02 or Later)
Follow the procedure below to replace the DRYPRO 832.
Requirements
Prepare following when replacing DRYPRO 832.
Note
This procedure is the replacing method when the software version of the DRYPRO 832 is V1.02 or later.
Refer to
“
Replacing DRYPRO 832 (Up to V1.01) (Page 17-82)”
if the software version is up to
V1.01.
For information about the software version of the DRYPRO 832, refer to
“15.11 System Setup (Page 15-
25)”
.
Caution
Do not use the compact flash from the malfunctioning DRYPRO 832 in the new DRYPRO 832
main unit.
If you accidentally replaced the compact flash (before turning ON the power)
Return to original compact flash before turning ON the DRYPRO 832.
When you have turned ON the power of DRYPRO 832
After checking following message on the operation panel of DRYPRO 832, turn OFF the power
breaker directly, and return the original compact flash.
•
“Which was changed? CF:enter MEC:reset”
Following message is displayed 3 minutes after above message is displayed.
•
“Turn OFF the power”
When above message is displayed, turn the power OFF of the DRYPRO 832 by pressing the
operation switch, then turn OFF the power breaker directly, and then return the original
compact flash.
When it has already started
After above message is completed, turn the power OFF of the DRYPRO 832 by pressing the
operation switch, then turn OFF the power breaker directly, and then return the original
compact flash.
•
“INITIALIZING”
Refer to operation manual for messages displayed on the operation panel.
Caution
If there are IP Addresses “192.168.20.161” or “192.168.20.179” used in the same network
environment of the facility, you need to release these IPaddresses before the procedure.
No.
Requirements
No.
Requirements
1
Replacement DRYPRO 832 main unit
(V1.02 or later)
2
Upgrading file
(V1.02 or later)
3
Backup file (DRYPRO 832)
4
DRYPRO 832 Operation Manual
5
DRYPRO 832 Installation Manual
6
DRYPRO 832 Repair Manual
START
Starting the server PC
If the power for the server PC is OFF, turn the power ON.
Caution
Connection error message is displayed when the server PC is CS-2 or CS-3.
Click the [Yes] button, and close the dialog box.
Summary of Contents for Drypro 832
Page 2: ......
Page 12: ......
Page 22: ......
Page 28: ......
Page 74: ......
Page 118: ......
Page 228: ......
Page 242: ......
Page 388: ......
Page 498: ......
Page 521: ...18 23 18 5 Block Diagram ...
Page 527: ...18 29 18 7 Printing Operation Load Timing Chart Standard 1 ch ...
Page 529: ......