Chapter 15 DRYPRO 832 Utility Tool
15-33
15.15 HPRO Setup
Check or change the constants for controlling the heat process unit.
1
Refer to
“15.1 DRYPRO 832 Utility Tool Start (Page 15-2)”
, and open the maintenance menu
window of the DRYPRO 832 Utility Tool.
2
Click the [HPRO SETUP] button under “SERVICE INFO”.
The “HPRO SETUP” window appears.
“HPRO SETUP” window - “HPRO” tab
The “HPRO” tab in the “HPRO SETUP” window contains the following items.
Caution
Each constant for HPRO setup is set to an optimum value at the factory. Do not change unless
otherwise instructed.
No.
Setup Items
Description of Settings
1
OFFSET (V1-V4)
The offset value for each item of the heat process temperature control is displayed.
Display range: -250 - 250
2
HEAT BASE (V2,
V4, V5 QC)
The temperature measured from each section of the heat process guide is displayed.