Chapter 15 DRYPRO 832 Utility Tool
15-26
3
Set each item according to the status of use in the facility.
4
Click the [SAVE] button.
Save the changes and close the “SYSTEM SETUP” window to return to the DRYPRO 832 Utility Tool Maintenance
Menu window.
If the network settings of “IMAGER NAME” were changed, restart the DRYPRO 832.
6
POWER OFF
MODE
Set the power OFF mode interfaced with Printlink5-IN.
Default value
:
OFF
OFF
The power OFF mode is not used.
ON (with
temperature
control)
The power OFF mode is used.
(Since the temperature inside the main unit is maintained even after
power is turned OFF, the warm-up period during the start-up
process will be approximately 3 minutes.)
ON (without
temperature
control)
The power OFF mode is used.
(Since the temperature control inside the main unit is not performed
after power is turned OFF, the warm-up period during the start-up
process will be approximately 10 minutes.)
For more information, see
“
About Power OFF Mode (Page 15-27)”
.
7
SLEEP MODE
TIMER
Set the sleep mode timer to ON/OFF when the DRYPRO 832 is not used, and set
the period of time before shifting into sleep mode.
Default value
:
OFF
ON
The sleep mode timer function is used.
OFF
The sleep mode timer function is not used.
sec
Enter the period of time before shifting into sleep mode. This value
can only be set when the sleep mode timer function is “ON”.
Default value:
120 min
8
BUZZER
Set the operation unit buzzer sound ON/OFF when an error or empty occurs in
DRYPRO 832.
Default value
:
ON
ON
The operation unit buzzer sounds for approximately 1 second.
OFF
The operation unit buzzer does not sound.
9
LOG WRITE
MODE
Set the log output status.
Default value
:
NORMAL
NORMAL
Output in normal mode.
DETAIL
Output in detail mode.
10
TOTAL
OPERATING TIME
(Display only)
The total operating time of DRYPRO 832 is shown.
Default value
:
0
11
TOTAL PRINTED
SHEETS
(Display only)
The total number of sheets output to DRYPRO 832 is shown.
The total number of printed sheets includes the number of sheets of film output
during calibration performed when the machine was installed.
Default value
:
0
12
VERSION
(Display only)
The currently active software version is displayed.
When the software is upgraded, confirm using this value.
Default value
:
Varies depending on shipment time.
13
[SAVE] button
Save the changes and close the “SYSTEM SETUP” window to return to the
DRYPRO 832 Utility Tool Maintenance Menu window.
14
[BACK] button
Discard the changes and close the “SYSTEM SETUP” window to return to
DRYPRO 832 Utility Tool Maintenance Menu window.
No.
Setup Items
Description of Settings
Summary of Contents for Drypro 832
Page 2: ......
Page 12: ......
Page 22: ......
Page 28: ......
Page 74: ......
Page 118: ......
Page 228: ......
Page 242: ......
Page 388: ......
Page 498: ......
Page 521: ...18 23 18 5 Block Diagram ...
Page 527: ...18 29 18 7 Printing Operation Load Timing Chart Standard 1 ch ...
Page 529: ......