Chapter 4 Replacing Supply Unit Loading Parts
4-13
Installation Procedure
1
Install the pickup unit assembly.
2
Install the light shield unit assembly.
• Two screws (M3
×
8)
3
Connect the connector (JJ48).
For Lispl-832, remove the Lispl-832 side connector (JJ48).
4
Install the lower front cover.
• Two screws (M3
×
8)
Caution
Insert the pickup unit assembly all the way in
along the guides while being careful not to
cause any damage by coming into contact
with the parts inside the main unit.
If the pickup unit assembly is not installed
properly, it may be impossible to install the
light shield unit assembly.
Caution
When installing the light shield unit
assembly, be careful not to cause the sensor
to come into contact with the parts inside the
main unit.
Caution
Be sure to push the connector all the way in
when connecting it.
If the connector is not connected properly,
an error will occur and, in some cases,
smoke and/or an odor may be generated.