Chapter 15 DRYPRO 832 Utility Tool
15-50
4
Click the [BACK] button.
The “DENSITOMETER ADJUST” window closes, returning to the Maintenance Menu window.
This completes the process of checking the set density value.
No.
Setup Items
Description of
Settings
Description of Settings
1
NORMAL FILM
INTERNAL
The results of measuring with a built-in densitometer are
shown, when a correction is made to BLUE film or CLEAR
film.
2
EXTERNAL
The results of measuring with an external densitometer are
shown, when a correction is made to BLUE film or CLEAR
film.
3
RESULT
If the result of comparing the internal density value with the
external density value is within
±
3, “OK” will be shown; or
if it is
±
4 or greater, ”NG” will be shown.
4
HIGH DENSITY
FILM
INTERNAL
The results of measuring with a built-in densitometer are
shown, when a correction is made to HIGH BLUE film.
5
EXTERNAL
The results of measuring with an external densitometer are
shown, when a correction is made to HIGH BLUE film.
6
RESULT
If the result of comparing the internal density value with the
external density value is within
±
3, “OK” will be shown; or
if it is
±
4 or greater, “NG” will be shown.