Chapter 17 Troubleshooting
17-96
Start the server PC
If the power for the server PC is OFF, turn the power ON.
Caution
Connection error message is displayed when the server PC is CS-2 or CS-3.
Click the [Yes] button, and close the dialog box.
Install the Printlink Build IN
(CS-2 and CS-3 only)
Refer to
“Chapter 7: Connecting and Setting CS-2 or CS-3” in the installation manual
to install
the latest Printlink Build IN.
However, select “BuiltIN” in the “Mode select” window at installation.
Upgrading server PC
(Printlink5-IN only)
Refer to
“16.4 Upgrade Information (Page 16-12)”
, and upgrade the
Printlink5 software to
a newer version than [V1.01R00].
Restore the setup
Only when the server PC is Printlink5-IN
Refer to
“16.6 Restoring Settings (Page 16-26)”
to restore the setup.
Caution
Perform the restore for the Printlink5-IN. Do not restore DRYPRO 832.
If you have restored DRYPRO 832, it will be set to setting from the previous
backup. Change the setting to the one with the changes made after previous
backup.
Restart the server PC
Restart the server PC (Printlink5-IN, CS-2, or CS-3)
Exit the application
(CS-2 and CS-3 only)
Refer to
“7.4 How to Exit CS-2 and CS-3” in the installation manual
to exit CS-2 or CS-3.
Start the DRYPRO 832
Refer to
“3.1.1 How to Start” in the operation manual
, and check whether the DRYPRO 832
starts normally.
Note
However, if the MEC upgrade is performed at startup, the startup time may take a few
minutes longer than normal.
Summary of Contents for Drypro 832
Page 2: ......
Page 12: ......
Page 22: ......
Page 28: ......
Page 74: ......
Page 118: ......
Page 228: ......
Page 242: ......
Page 388: ......
Page 498: ......
Page 521: ...18 23 18 5 Block Diagram ...
Page 527: ...18 29 18 7 Printing Operation Load Timing Chart Standard 1 ch ...
Page 529: ......