Chapter 15 DRYPRO 832 Utility Tool
15-43
4
Choose the tray for which you want to perform densitometer calibration, and then click the
[PRINT] button.
Example: Calibration pattern
One sheet of calibration pattern is printed.
Caution
Precautions when the optional Lispl-832 is installed:
•
If the same film type is used for Tray 1 and Tray 2, by performing the densitometer calibration
once and specifying one tray, calibration is performed for the other tray as well. If different
sizes are used, select the tray into which larger film is loaded.
•
If the film type in Tray 1 and Tray 2 are different, select the tray according to the following
table.
Combination of film types
Target tray for densitometer calibration
BLUE
film
CLEAR
film
Tray into which BLUE film is loaded
BLUE
film
HIGH BLUE
film
Tray into which BLUE film and HIGH BLUE film are loaded (*)
CLEAR
film
HIGH BLUE
film
Tray into which CLEAR film and HIGH BLUE film are
loaded (*)
(*) Perform density calibration for each tray.