Chapter 17 Troubleshooting
17-79
When Mistake was Made Selecting Replaced Unit
When mistake was made selecting replaced unit, reset necessary items by following procedure.
START
Starting the server PC
If the power for the server PC is OFF, turn the power ON.
Caution
Connection error message is displayed when the server PC is CS-2 or CS-3.
Click the [Yes] button, and close the dialog box.
Exit the application
(Only for CS-2 and CS-3)
Refer to
“7.4 How to Exit CS-2 and CS-3” in the installation manual
to exit CS-2 or CS-3.
Start the DRYPRO 832
Refer to
“3.1.1 How to Start” in the operation manual
, and check whether the DRYPRO 832
starts normally.
Note
However, if the MEC upgrade is performed at startup, the startup time may take a few
minutes longer than normal.
Initializing network
settings
Return the Printlink5-IN network settings to the factory settings.
This item does not need to be set if DRYPRO 832 is used with default network
setting (IP Address: 192.168.20.161)
Caution
Check that the following address is not the same as another IP address within
the facility.
•
IP Address: 192.168.20.179
•
Subnet mask: 255.255.255.0
1. Select [Start] – [Control Panel] – [Network Connections] from the WindowsXP
desktop.
The network connections window opens.
2. Right-click on “Local Area Connection”, and select “Properties” from the pull down
menu.
“Local Area Connection Properties” dialog is displayed.
Summary of Contents for Drypro 832