Chapter 17 Troubleshooting
17-73
Replacing Compact Flash (V1.02 or Later)
Follow the procedure below to replace the compact flash.
Requirements
Prepare following when replacing compact flash (CF).
Note
This procedure is the replacing method when the software version of the DRYPRO 832 is V1.02 or later.
Refer to
“
Replacing Compact Flash (Up to V1.01) (Page 17-64)”
if the software version is up to
V1.01.
For information about the software version of the DRYPRO 832, refer to
“15.11 System Setup (Page 15-
25)”
.
Caution
•
Before replacing the compact flash, replace the main board as shown in
“17.3.1 Replacing
Main Board (Page 17-63)”
, and check that the DRYPRO 832 main unit starts up normally.
•
If there are IP Addresses “192.168.20.161” or “192.168.20.179” used in the same network
environment of the facility, you need to release these IP addresses before the procedure.
•
Data for “Film Count” and/or “Maintenance Schedule” may be lost if the procedure is not
carried out correctly. Loaded film needs to be discarded and newly loaded when the data is
lost, so take note of the “Film Count” and “Maintenance Schedule” before starting the
procedure.
For details, refer to
“15.9 Film Setup (Page 15-20)”
,
“15.14 Maintenance Schedule (Page 15-
31)”
, and
“15.9 Film Setup (Page 15-20)”
.
No.
Requirements
No.
Requirements
1
Replacement compact flash
(V1.02 or later)
2
Upgrading file
(V1.02 or later)
3
Backup file (DRYPRO 832)
4
DRYPRO 832 Operation Manual
5
DRYPRO 832 Installation Manual
6
DRYPRO 832 Repair Manual
START
Exit the DRYPRO 832
unit
Refer to
“3.2.1 How to Exit” in the operation manual
to exit the DRYPRO 832 and switch the
Power source breaker OFF.
Replacing the compact
flash
Refer to
“11.4 Replacing Compact Flash (Page 11-11)”
to replace the compact flash.
Starting the server PC
If the power for the server PC is OFF, turn the power ON.
Caution
Connection error message is displayed when the server PC is CS-2 or CS-3.
Click the [Yes] button, and close the dialog box.
Exit the application
(Only for CS2 and CS-3)
Refer to
“7.4 How to Exit CS-2 and CS-3” in the installation manual
to exit CS-2 or CS-3.
Upgrade server PC
Refer to
“16.4 Upgrade Information (Page 16-12)”
, and upgrade the version of server PC to
V1.02R00 or later.
If the installation completes normally, the server PC will automatically restart.
Caution
Connection error message is displayed when the server PC is CS-2 or CS-3.
Click the [Yes] button, and close the dialog box.
Summary of Contents for Drypro 832
Page 2: ......
Page 12: ......
Page 22: ......
Page 28: ......
Page 74: ......
Page 118: ......
Page 228: ......
Page 242: ......
Page 388: ......
Page 498: ......
Page 521: ...18 23 18 5 Block Diagram ...
Page 527: ...18 29 18 7 Printing Operation Load Timing Chart Standard 1 ch ...
Page 529: ......