Chapter 4 Replacing Supply Unit Loading Parts
4-12
4
Remove the lower front cover.
• Two screws (M3
×
8)
5
Remove the connector (JJ48).
For Lispl-832, remove the Lispl-832 side connector (JJ48).
6
Remove the light shield unit.
• Two screws (M3
×
8)
7
Remove the pickup unit assembly.
This is the end of the process of removing the pickup unit assembly.
•
When replacing the pickup transport motor, go to
“4.2.3 Replacing Pickup Transport Motor (Page 4-15)”
.
•
When replacing the pickup HP motor, go to
“4.2.4 Replacing Pickup HP Motor (Page 4-18)”
.
•
When replacing the pickup HP sensor, go to
“4.2.5 Replacing Pickup HP Sensor (Page 4-20)”
.
Caution
When removing the connector, do not pull by
the wires, but hold the tab on the connector
with your fingers.
Pulling by the wires too tightly can cause a
break.
Caution
When removing the light shield unit
assembly, be careful not to cause the sensor
to come into contact with the parts inside the
main unit.
Caution
When removing the pickup unit assembly,
be careful not to cause the sensor to come
into contact with the parts inside the main
unit.