Chapter 15 DRYPRO 832 Utility Tool
15-34
“HPRO SETUP” window - “MEC” tab
The “MEC” tab in the “HPRO SETUP” window contains the following items.
3
NORMAL
(V1-V4)
The set temperature of the heat process guide (V1-V4) under normal conditions is
displayed.
Display range: 900-1400
(
0.1
℃
units
)
Default: V1
1000
(
900 - 1100
)
Default: V2
1265
(
1200- 1400
)
Default: V3
1265
(
1200 - 1400
)
Default: V4
1000
(
900 - 1100
)
4
SLEEP MODE
(V1-V4)
The set temperature of the heat process guide (V1-V4) in sleep mode is displayed.
Display range: 1200-1300
(
0.1
℃
units)
Default: V1
750
(
700 - 1100
)
Default: V2
1250
(
1000 - 1250
)
Default: V3
1250
(
1000 - 1250
)
Default: V4
850
(
700 - 1000
)
5
SETUP DATA
The parameter used for density correction control is displayed.
Display range: 0-300
6
[SAVE] button
Save the changes and close the “HPRO SETUP” to return to DRYPRO 832 Utility
Tool Maintenance Menu window.
7
[BACK] button
Discard the changes and close the “HPRO SETUP” to return to DRYPRO 832
Utility Tool Maintenance Menu window.
No.
Setup Items
Description of Settings
1
OFFSET (V1-V4)
The offset value for each item of the heat process temperature control is displayed.
Display range: -250 - 250
No.
Setup Items
Description of Settings
Summary of Contents for Drypro 832
Page 2: ......
Page 12: ......
Page 22: ......
Page 28: ......
Page 74: ......
Page 118: ......
Page 228: ......
Page 242: ......
Page 388: ......
Page 498: ......
Page 521: ...18 23 18 5 Block Diagram ...
Page 527: ...18 29 18 7 Printing Operation Load Timing Chart Standard 1 ch ...
Page 529: ......