Chapter 5 Replacing Tray Unit Loading Parts
5-18
11
Remove the tray lock solenoid unit assembly.
• Three screws (M3
×
6)
The bracket is hooked on the notch in the main unit, but can
be unhooked by sliding slightly to the right and pulling back.
12
Push the tab on the back of the sensor inward
and remove the photo sensor.
13
Insert a new photo sensor in the mounting hole.
14
Install the tray lock solenoid unit assembly
removed in
step 11
.
• Three screws (M3
×
6)
After inserting the bracket notch into the notch groove of the
main unit ( in figure on left), align the position of the other
notches and attach.
15
Connect the connectors removed in
step 10
.
•
If the tray unit tray detection sensor has been replaced, connect the connector (JP55, JP44).
•
If the Lispl-832 tray detection sensor has been replaced, connect the connector (JP63, JP44).
16
Install the connector cover removed in
step 7
or
step 9
.
• One screw (M3
×
6)
Caution
Be sure to push the connector all the way in when connecting it.
If the connector is not connected properly, an error will occur and, in some cases, smoke and/or
an odor may be generated.
Note
If connecting the Lispl-832, connect connectors (JP16, JP 26).
Note
•
If the tray unit tray detection sensor has been replaced, proceed to
step 17
.
•
If the Lispl-832 tray detection sensor has been replaced, proceed to
step 18
.
Summary of Contents for Drypro 832
Page 2: ......
Page 12: ......
Page 22: ......
Page 28: ......
Page 74: ......
Page 118: ......
Page 228: ......
Page 242: ......
Page 388: ......
Page 498: ......
Page 521: ...18 23 18 5 Block Diagram ...
Page 527: ...18 29 18 7 Printing Operation Load Timing Chart Standard 1 ch ...
Page 529: ......