Chapter 15 DRYPRO 832 Utility Tool
15-45
7
Measure the density of the printed densitometer
calibration patterns (01) - (12), using an external
densitometer.
8
Click the [Next(N)] button.
The “DENSITOMETER(STEP3)” window appears.
9
Enter a value equals to the density measured in
Step 7
, which was increased by 100-fold, in
one-byte numbers in [01] - [12] in the “DENSITOMETER (STEP3)” window,
Example:
•
If the (01)st density value is 0.21, enter “21” in the 01 field.
•
If the (12)th density value is 3.16, enter “316” in the 12 field.
Caution
When measuring the film density, make sure
to standardize the measuring surface of film
(front/back side).
If the measuring surface is not standardized,
the results of measuring the film density may
vary, depending on the model of the
densitometer.
Note
For details of how to measure the density, see the
instruction manual that comes with the
densitometer.
Caution
Entering and saving an abnormal or incorrect value can cause density correction data
corruption, making it impossible to perform density correction properly. Take care not to enter
an abnormal or incorrect value.
•
Example of abnormal or incorrect values
•
Values deviating greatly from the measured value
•
The density value was 1.25, but “215” was entered by mistake.
•
The density value was 1.25, but an abnormal value of “400” was intentionally entered.
Summary of Contents for Drypro 832
Page 2: ......
Page 12: ......
Page 22: ......
Page 28: ......
Page 74: ......
Page 118: ......
Page 228: ......
Page 242: ......
Page 388: ......
Page 498: ......
Page 521: ...18 23 18 5 Block Diagram ...
Page 527: ...18 29 18 7 Printing Operation Load Timing Chart Standard 1 ch ...
Page 529: ......