Capter 3
Function Setting and Basic Operation
Curve Tracer CS-3000 Series Instruction Manual
3-83
■
If message "This option has already been installed." appears on lower side of screen
The message may appear on OPTION INSTALL
STATUS screen as shown on the right.
This indicates
that the corresponding option is already installed
in the instrument.
Since the option saved in USB memory can be used,
it is not necessary to install it. Remove USB memory.
■
If error message "This file cannot be installed in this machine.” appears on lower side of
screen
When the error message appears in OPTION INSTALL
STATUS
screen as shown on the right, possible causes are:
- Product ID of the optional file saved in USB
memory is different from that of this instrument.
- The optional file saved in USB memory is not
correct.
<Action method>
①
Check Product ID to confirm if the option is
correct to the instrument in which the option is
to be installed. If a wrong USB is inserted, use
USB memory consistent with Product ID of this
instrument.
②
Execute [Procedure] of installation again.
If the error message appears again,
contact Iwatsu office or our sales distribution.