Capter 3
Function Setting and Basic Operation
3-44
Curve Tracer CS-3000 Series Instruction Manual
3.9.2 VERTICAL knob
Rotating VERTICAL knob allows the sense and source of the vertical axis to be selected.
The setting result is displayed in the setup display area under the waveform display area.
50.0 mA
Vertical axis range
COLLECTOR
Source
Setting range
Setting ranges in HIGH CURRENT mode depend on models and the maximum peak power
setting.
Source: COLLECTOR
If other than Open/Collector/Common is selected on CONFIGURATION menu, COLLECTOR
current (Ic) is measured. If Open/Collector/Common is set, BASE current (Ib) is measured when
the source is set to COLLECTOR.
VERTICAL
←
To left rotation
To right rotation
→
Source
STEP
COLLECTOR
Range
Fixed
position
display
HIGH VOLTAGE:
CS-3100, CS-3200,
CS-3300
2.00A/div to 1.00uA/div 1-2-5 step
HIGH CURRENT:
CS-3300
10kW(1kA) : 100A/div to 10.0A/div 1-2-5 step
4kW(400A) : 100A/div to 1.00A/div 1-2-5 step
400W(40A) : 100A/div to 100mA/div 1-2-5 step
HIGH CURRENT:
CS-3200
10kW(1kA) : 50.0A/div to 10.0A/div 1-2-5 step
4kW(400A) : 50.0A/div to 1.00A/div 1-2-5 step
400W(40A) : 50.0A/div to 100mA/div 1-2-5 step