Capter 3
Function Setting and Basic Operation
Curve Tracer CS-3000 Series Instruction Manual
3-73
< The second page of MEASUREMENT SETUP submenu >
NO.
Submenu
Function description
F1
VARIABLE KNOB RESPONSE
The VARIABLE knob sets the output voltage of the collector supply
at the rate to the maximum peak voltage setting value. When the
VARIABLE knob is turned, the collector supply voltage can be
continuously set from 0 % to 100 % in 0.1% step
The response (change rate) when the VARIABLE knob is turned
can be set by two stages of the following.
NORMAL: Usual response (change rate)
QUICK
: The response (change rate) quickens.
F2
LEAKAGE MEASUREMENT
When the MODE setting of the collector supply is set to LEAKAGE
and the current of the emitter is measured, the measurement mode
can be selected from the following.
-
AVERAGE:
When measuring it, the leveling processing is done
-
POINT:
When measuring it, the leveling processing is not done, and
the measurement value is sampled every intervals of time.
F3
~
F5
(Not used)
F6
PREV PAGE
The menu is switched to PREV PAGE by pushing F6 key.