Capter 3
Function Setting and Basic Operation
3-92
Curve Tracer CS-3000 Series Instruction Manual
3.11.4 Setting of SAVE/RECALL/DELETE (saving/recalling/deleting data)
This section explains procedures for SAVE/RECALL/DELETE for each data type, and internal
memory/USB memory.
1. Data type
Data types to be handled by SAVE/RECALL menu are as follows:
- SETUP: Setting condition data
- TRACE: Setting condition data and characteristic curve data
- REF: REF characteristic curve data
2. Internal memory/USB memory
Setting for SAVE uses the submenu [TO] and setting for RECALL and DELETE uses the
submenu [FROM].
- Internal memory in this instrument
- USB memory
3. File name (when saving into USB memory)
- File name (Ex.: ABCD1234) consists of 8 characters; the first 4 alphabet characters (Alphanumeric
character of normal-width capital letter;!#$%&'()-@^_{}~) and 4 numeric 8-bit characters following
them.
- When saved into the USB memory, data is saved with the specified file name in the SETUP, WAVE,
or REF folder. If folder above does not exist, the folder is automatically created.
[Procedure]
(1) Each time data is saved into the USB memory, 4-digit number on the right in the file name increases
one by one. (Note that when reaching 9999, the number is not increased.)
(2) The folders and extensions below are used for each type of data being saved.
a) SETUP
¥SETUP¥*.SET
b) TRACE
¥WAVE¥*.WFM: binary format
¥WAVE¥*.CSV: text format
c) REF
¥REF¥*.REF
(3) If there is the file with the same file name when saving the file, the confirmation screen to overwrite is
displayed.
(4) Pressing OK key allows the file to be overwritten.
Caution!
♦
Execution of RECALL causes the voltage being set by VARIABLE % to be applied to a measured
device. Care should be taken.