6.1 Instrument specifications
Hereinafter, HIGH VOLTAGE will be expressed as HV and HIGH CURRENT as HC
(1/4)
Items
Instrument specifications
Collector supply
(HV mode)
Mode/polarity
AC
±
full-wave rectified
±
DC
±
LEAKAGE
Max. peak voltage
30 V, 300 V, 3 kV (AC: 2.5 kV)
Max. peak current
/pulse peak current
Max. peak voltage Max. peak current Pulse peak current
30 V
7.5 A
15 A
300 V
750 mA
1.5 A
3k V
75 mA
150 mA
Series resistance
Max. peak voltage
Max. peak power
30 V
300 V
3 kV
390 W
200 m
Ω
20
Ω
-
120 W
2
Ω
200
Ω
20 k
Ω
12 W
20
Ω
2 k
Ω
200 k
Ω
1.2 W
200
Ω
20 k
Ω
2 M
Ω
120 mW
2 k
Ω
200 k
Ω
20 M
Ω
Max. peak power
120 mW, 1.2 W, 12 W, 120 W, 390 W
Note:
If Max. peak voltage is set to 3 kV, 390 W cannot be used.
Variable collector supply
Continuously variable from 0
%
to 100.0% (in 0.1
%
step)
Note: Output not calibrated
Output enabled/disabled
Collector supply can be set to output enabled/disabled.
Loop calibration
Hardware
Floating capacity between collector supply and ground can be
calibrated.
Software
Software thinning pseudo-calibrates the loop.
Collector supply
(HC mode)
∗
Specifications for
CS-3200/3300
Mode/polarity
±
PULSE
Max. peak voltage
40 V
Max. peak current
Max. peak power
Max. peak current
10 kW
1 kA (only CS-3300)
4 kW
400 A
400 W
40 A
Output enabled/disabled
HC mode allows output from the collector supply to be
ENABLEd or DISABLEd.
Pulse width
50 us to 400 us (in 10 us step)
Note: If the pulse width is narrow, up to max. peak voltage may
not be generated.
Pulse interval
80 ms to 1000 ms (a 400 W)
160 ms to 1000 ms (at 4 Kw)
320 ms to 1000 ms (at 10 kW)
Pulse measurement point
10 us to Pulse width setting value (in 10 us step)
No. of data pieces
20 to 1000 points/trace