Capter 3
Function Setting and Basic Operation
Curve Tracer CS-3000 Series Instruction Manual
3-35
TYPE can be selected. Refer to section 3.11.7 for details of this option CS-801.
High Current mode (This mode is available on CS-3200 and CS-3300.)
Key/knob
ACQUIRE
Function description
F1
FUNCTION
PULSE INTERVAL
Sets the interval of collector output pulse.
The possible ranges to be set depend on the maximum peak power setting
(as shown in table below).
F2
FUNCTION
SWEEP STEPS
Available when HORIZONTAL source is COLLECTOR.
It sets the number of measurement points per step at the time of collector
supply SWEEP
Setting range: 20 to 1000 (increment of 1 STEP)
F3
SWEEP TYPE
Selects SWEEP direction from DOWN/UP/CUSTOM/DOWN UP.
Note
.
- DOWN:
It sweeps the range from VARIABLE setting value
when SWEEP starts to 0.0%.
- UP:
It sweeps the range from 0.0% to VARIABLE setting
value when SWEEP starts.
- CUSTOM:
It sweeps the range from VARIABLE setting value
when SWEEP starts to the setting value of SWEEP
END VARIABLE.
- DOWN UP:
Note
Sweep from the VARIABLE set value in point that
starts SWEEP to the set value of SWEEP END
VARIA BLE. In addition, turn as it is continuously, and
sweep up to the VARIABLE set value in point that
starts SWEEP in the opposite direction.
F4
FUNCTION
SWEEP END
VARIABLE
Available only when SWEEP TYPE is CUSTOM/DOWN UP
Note
.
It sets VARIABLE setting value to finish or turn SWEEP.
Setting range: 0 to 100
%
(increment of 0.1
%
)
F5
SET CURRENT
VARIABLE
Available only when SWEEP TYPE is CUSTOM/DOWN UP
Note
.
Pressing F5 key allows the current VARIABLE setting value to be set to
SWEEP END VARIABLE.
F6
FUNCTION
SWEEP WAIT TIME
It can set, in SWEEP operation, the waiting time after a voltage changes in
VARIABLE until next data is acquired. (It is available when, e.g.,
measurement considers the influence of time constant on a measured
device.)
Note) When the use of DOUBLE SWEEP optional CS-801 and this function have been turned on, DOWN UP of
SWEEP TYPE can be selected. Refer to section 3.11.7 for details of this option CS-801.
Ranges for PULSE INTERVAL setting in HIGH CURRENT mode are as follows depending on the
maximum peak power setting. In both of REPEAT measurement and SWEEP measurement,
pulses are generated with this cycle for measurement.
Conditions for collector supply SWEEP
- The collector supply SWEEP function in HIGH VOLTAGE mode is available if the step generator is set to
PULSE and HORIZONTAL source is set to COLLECTOR.
- SWEEP function in HIGH CURRENT mode is available if HORIZONTAL source is set to COLLECTOR.
Maximum peak power setting
PULSE INTERVAL setting ranges
10kW
320ms to 1000ms
4kW
160ms to 1000ms
400W
80ms to 1000ms