Chapter 6 Specifications
Curve Tracer CS-3000 Series Instruction Manual
6-3
(3/4)
Items
Instrument specifications
Horizontal axis
Collector voltage
Range
HV:50 mV/div to 500 V/div, 13 steps, 1-2-5 switching
HC:50 mV/div to 5 V/div , 7 steps, 1-2-5 switching
Accuracy 2
%
of readout
+
0.05
×
HORIZ/div setting value or less
Base/emitter voltage
Range
Accuracy
50 mV/div to 5V/div, 7 steps, 1-2-5 switching
2
%
of readout
+
0.05
×
HORIZ/div setting value or less
Step generator display
Range 0.5 step/div
POSITION
Inside of tube: 10 div
Magnification display
×
2,
×
5,
×
10
Display
Type
8.4-inch TFT-LCD
Resolution
SVGA:800
×
600 pixel
No. of data pieces
1000 points/trace (AC full-wave rectified)
20 to 1000 points/trace (SWEEP mode)
Trace display
Between-point interpolation display, dot display
Average
Off, 2 to 255 times
Persistence
Off, Short, Long, infinite
Internal waveform storage
(REF)
4 screens can be saved.
Cursor measurement
DOT cursor
Vert, Horiz,
β
, or gm
fLINE cursor
FREE cursor
Vert, Horiz, 1/grad, intercept
Vert, Horiz ,
β
, or gm
WINDOW cursor
Vert, Horiz,
β
, or gm in WINDOW area
Built-in timer
Display
Year/month/day/hour/minute
Internal memory
Setup memory
256 memories
External memory
USB memory
USB1.1
Setup, waveform save/recall, screen hard copy
Interface
10 Base-T100 Base-TX Ethernet
Power supply
Power input range
AC 100-240 V, 50/60 Hz
Power consumption(Operating) 500 VA (400 W) Max
Power consumption(Standby)
50 VA
(
7 W
)
Max
Summary of Contents for CS-3000 Series
Page 1: ...Instruction Manual Curve Tracer CS 3000 series CS 3100 CS 3200 CS 3300 ...
Page 2: ...Ⓒ 2010 IWATSU ELECTRIC CO LTD All rights reserved ...
Page 16: ......
Page 170: ......