Capter 3
Function Setting and Basic Operation
3-18
Curve Tracer CS-3000 Series Instruction Manual
Names and displayed contents
NO.
Name
Displayed content
Reference
①
State display
The output state of the collector supply is displayed in 3 statuses
below:
INTERLOCK:
It is displayed when the test box is not connected
correctly or the cover of the test fixture is opened.
DISABLE:
It shows that the collector supply can not output.
There are two kinds of breaker switches for
COLLECTOR SUPPLY about HIGH VOLTAGE and
HIGH CURRENT.
When the HIGH VOLTAGE breaker switch is
adjusted to the DISABLED side for the HIGH
VOLTAGE mode, it is displayed on the left of the
screen, "DISABLE". When the HIGH CURRENT
breaker switch is adjusted to the DISABLED side for
the HIGH CURRENT mode, it is displayed on the left
of the screen, "DISABLE".
ENABLE:
It shows that the collector supply can output.
When the lid
of test fixture is shut on the ENABLED
side, the breaker switch for HIGH VOLTAGE is
displayed for the HIGH VOLTAGE mode, "ENABLE".
When the lid of test fixture is shut on the ENABLED
side, the breaker switch for HIGH CURRENT is
displayed for the HIGH CURRENT mode, ENABLE".
ENABLE is set when the breaker switch for
COLLECTOR SUPPLY is set to ENABLED and the
cover of the test fixture is closed.
Even in ENABLE status, when measurement stops
(i.e., measurement mode: STOP), the collector supply
does not output. The actual status of the collector
supply output can be checked by the icons as shown
below. Note that in LEAKAGE mode, even when
measurement stops, the collector supply continues to
output.
When COLLECTOR SUPPLY output is ON
When COLLECTOR SUPPLY output is OFF.
Section 3.3.2
②
Various messages
Displays execution state of the menu and error massages
③
Measurement mode
Displays the measurement mode of REPEAT/SINGLE/STOP/SWEEP
Section 3.8
④
Number of averages
Displays the first progress state. In REPEAT mode, it displays each
result after that. The number of averages is 2 to 255.
Section 3.10.1
⑤
Header information
Displays the current time, CONFIGURATION setting, or comment as
the header information.
Section 3.11
⑥
Function menus
Pressing a function key (F1 to F6) allows a submenu to appear.
Section 3.5
⑦
Measurement result,
or waveform source
Displays the cursor measurement result, the waveform source for
WAVE display, and the range settings of vertical, horizontal and time
base.
Section 3.10.3
Section 3.10.4
⑧
COLLECTOR
SUPPLY
Displays MODE, polarity, VARIABLE, Watt, voltage, and limitation
resistance.
Section 3.7
⑨
STEP GENERATOR Displays MODE, range, number of steps, and offset value.
Section 3.6
⑩
HORIZONTAL
Displays the horizontal axis range and source.
Section 3.9.1
⑪
VERTICAL
Displays the vertical axis range and source.
Section 3.9.2