Capter 3
Function Setting and Basic Operation
3-32
Curve Tracer CS-3000 Series Instruction Manual
3.7.3 POWER
−
WATT key
3.7.3.1
POWER-WATT key and display menu
It can limit the maximum peak power to be applied to a measured device; i.e. protecting the
device from breakage.
Pressing POWER - WATT key
③
allows "MAX PEAK POWER WATTS" menu to appear.
High Voltage mode
High Current mode
3.7.3.2
Submenus and functions
This setting value and the maximum peak voltage setting value determine the value of the
current limitation resistance internally connected to the collector output in series. See Table 3.7
POWER WATTS/VOLTAGE and the current limitation resistance in Section 3.7.
In addition, if the maximum peak voltage setting is 3 kV, 390 W cannot be selected.
High Voltage mode
High Current mode
F1
Sets the maximum power to be applied to
a measured device (390W, 120W, 12W,
1.2W, or 120mW).
Sets the maximum power (current) to be applied to
a measured device (10kW, 4kW, or 400W).
F2
F3
F4
(Not used)
F5
F6
(Not used)
* CS-3200 and CS-3300 include
HIGH CURRENT.
F1
F2
F3
F4
F5
F6
Summary of Contents for CS-3000 Series
Page 1: ...Instruction Manual Curve Tracer CS 3000 series CS 3100 CS 3200 CS 3300 ...
Page 2: ...Ⓒ 2010 IWATSU ELECTRIC CO LTD All rights reserved ...
Page 16: ......
Page 170: ......