Capter 3
Function Setting and Basic Operation
Curve Tracer CS-3000 Series Instruction Manual
3-55
3.10.4.2
fLINE cursor
It displays the straight line cursor which passes on DOT cursor specified by STEP and position
(index).
Rotating FUNCTION knob allows the gradient to change.
Values of vertical axis (VERT: mA) and horizontal axis (HORIZ: V), the step number, and the
index position of the cursor position being set by DOT cursor are displayed in the area under
the menu.
The gradient of fLINE cursor is displayed under it.
Gradient:
- When the horizontal axis is used for COLLECTOR voltage: Vc/Ic
=
resistance (
Ω
) (1/grad display)
- When the horizontal axis is used for BASE-EMITTER voltage: Ic/Vbe
=
trans conductance (mS) (grad
display).
In addition, the voltage value at the intersection point of fLINE cursor and the horizontal axis is
displayed in the intercept readout. If the horizontal or vertical axis is used for STEP, no cursor
appears and the parameter above is not calculated.