ii
Read this page to ensure proper safety.
(Also read the following pages.)
WARNING
● In measurement with this instrument, high voltage may be
applied and high current may be supplied. Always close the
cover before measurement after setting the DUT (Device
Under Test) on the test fixture.
High voltage (max. 3kV) may be applied and high current (max. 1000A) may be supplied to the
terminal to which the DUT is mounted. To prevent danger, if the cover of the test fixture is not closed,
no voltage and current can be supplied to the terminal and no power can be applied to the DUT. And,
do not remodel the switch for power supply shutoff or do not open the cover during measurement. If
not, electric shock may be caused.
● Do not open the cover of the test fixture during
measurement.
High voltage (max. 3kV) and high current (max. 1000A) may be applied to the terminal to which the
DUT is mounted. To prevent danger or to protect the DUT, when detecting that the cover of the test
fixture opens, the instrument immediately shuts off the voltage and current applied to the terminal and
stops the power supply to the DUT. And for safety, do not open the cover of the test fixture during
measurement.
● After measurement, the DUT may be charged. Do not touch
the DUT until discharging it through the grounding rod and
confirming the safety.
Touching the charged DUT may cause electric shock.
● Do not use other than attached Test Lead sets
(alligator-banana clip)
(Only for CS-3200/CS-3300)
If measurement is done without considering measurement electric potential and usage rating of the cable
outer cover and GND cable outer cover, fire or failures may be caused. Use the specified cables
attached depending on measurement types.