Capter 3
Function Setting and Basic Operation
3-112
Curve Tracer CS-3000 Series Instruction Manual
COLLECTOR SUPPLY SETUP submenu
Pressing F1 key allows COLLECTOR SUPPLY SETUP submenu to appear.
Key/knob
Setting items
Function description
F1
-
Displays the polarity of the output form COLLECTOR SUPPLY.
It depends on the polarity of MODE/POLARITY menu of COLLECTOR
F2
FUNCTION
MINIMUM
Sets the minimum value of SWEEP range of COLLECTOR SUPPLY.
If any value larger than MAXIMUM below or START VARIABLE setting
value is set, the value of MAXIMUM or START VARIABLE is also
changed to the same value
- Setting range: 0.0 to 100.0%
- Setting resolution: 0.1%
F3
FUNCTION
MAXIMUM
Sets the maximum value of SWEEP range of COLLECTOR SUPPLY.
If any value smaller than MINIMUM or START VARIABLE setting value
is set, the value of MINIMUM or START VARIABLE is also changed to
the same value.
- Setting range: 0.0 to 100.0%
- Setting resolution: 0.1%
F4
FUNCTION
RESOLUTION
Sets the SWEEP resolution of COLLECTOR SUPPLY. If this option is
used for operation, the resolution in this setting is used for SWEEP.
Setting range: 0.1
~
5.0
%
- Setting resolution: 0.1%
F5
FUNCTION
START
VARIABLE
Sets START value of SWEEP of COLLECTOR SUPPLY.
The value should be set between values of MAXIMUM and MINIMUM
If any value out of the range from MAXIMUM to MINIMUM is to be set,
the value is set to MAXIMUM or MINIMUM value; i.e., any value out of
the range cannot be set..
- Setting range: 0.0 to 100.0%
- Setting resolution: 0.1%
F6
-
(Not used)
F1
F2
F3
F4
F5
F6