Chapter 4 Characteristics Measurement
4-20
Curve Tracer CS-3000 Series Instruction Manual
■
Measurement procedure
After step (19) in Section 4.1.2.1, I
C
vs. V
CE
characteristics is displayed.
(1) Rotate HORIZONTAL range counterclockwise and set to STEP.
(2) Press STEP GENERATOR MENU key.
(3) Rotate FUNCTION knob to set NUMBER OF STEPS to 10.
The characteristics of the collector current vs. the base current are displayed with one
horizontal division equal to one step.