Capter 3
Function Setting and Basic Operation
Curve Tracer CS-3000 Series Instruction Manual
3-21
3.6.1 Setting STEP knob
3.6.1.1
STEP knob and display menu
Rotating STEP knob
①
allows the voltage or current per step to be set.
Setting range
When STEP 1/10 key is set to OFF
STEP knob
SOURCE
←
To left rotation
To right rotation
→
VOLTAGE
50mV, 100mV, 200mV, 500mV, 1V, 2V
CURRENT
50nA, 100nA, 200nA, 500nA, 1μA, 2μA, 5μA, 10μA, 20μA, 50μA, 100μA,
200μA, 500μA, 1mA, 2mA, 5mA, 10mA, 20mA, 50mA, 100mA, 200mA
The menu above shows the setting values.