Capter 3
Function Setting and Basic Operation
Curve Tracer CS-3000 Series Instruction Manual
3-51
F4 and F5 settings
Settings of F4 (PULSE WIDTH) and F5 (MEASUREMENT POINT) is enabled only when the
collector supply is set to HIGH CURRENT or DC and the step generator is set to PULSE waveform.
When the collector supply is set to AC, RECTIFIED SINE (full-wave rectified), or LEAKAGE and
the step generator is set to STEP waveform, the settings are disabled. However, it is available if
the step generator is set to PULSE even if the collector supply is set to AC.
To set the width of pulses to be generated from the step generator, press F4 key and rotate
FUNCTION knob.
To set the MEASUREMENT POINT which is adopted as the measurement value (common to
HORIZONTAL and VERTICAL), press F5 key and rotate FUNCTION knob.
Step generator in PULSE setting
When VIEW MODE
=
WAVE, the cursor
indicating the point which is adopted as the
measurement value is displayed on the
waveform screen.
Full scale time axis in WAVE mode
It varies depending on selection between STEP
and PULSE of the step generator as shown
below:
Step generator =PULSE
Full scale display 500μs (50μs/div)
Step generator =STEP
STEP GENERATOR
STEP
PULSE
F4
Disable
Sets the pulse width (50us to 400us, in 10us step).
When the collector supply is in HGH CURRENT mode, the width of the output
pulse from the step generator is wider than that of the output pulse from the
collector supply; i.e., additional 100 us at each end.
F5
Sets the measurement timing (from 10us to pulse width setting value(F4), in 10us
step, (Max. value: 400 us)
F1
F2
F3
F4
F5