Capter 2
Name and Function of Each Part (Outline)
2-16
Curve Tracer CS-3000 Series Instruction Manual
Caution!
-
Do not apply a voltage exceeding 1 kV or a current exceeding 1 A to each terminal of the
test adapter (TO TYPE ADAPTER) CS-501A.
If 1 kV or 1 A is exceeded, test adapter or a measured device might be damaged.
- For an accurate measurement, insert an accessory insulation tab in the socket (
②
or
③
of Figure 2.13) that doesn't connect the measurement element while measuring.
Memo
- It is not possible to measure at the same time though two kinds of devices to be measured
can be installed in test adaptor CS-501A. Measure by installing one device to be
measured and one insulation tab.
Table 2.12 Name and function of each part on test adapter CS-501A (TO type adapter)
No.
Name
Description
Ref.
①
Base resistance
It certainly connects either the resistance element or the
short-circuit terminal.
Section 3.3.5
②
Sockets for
element: B C E
It connects the element arranged in the order of the base,
collector, and emitter. The groove in the center connects the
terminal of a measured element or the insulation tab.
Section 3.3.5
③
Sockets for
element: C B E
It connects the element arranged in the order of the collector,
base, and emitter. The groove in the center connects the terminal
of a measured element or the insulation tab.
Section 3.3.5
④
⑦
Collector
Collector sense
It connects with the collector (C) of the socket of
①
or
②
in
Figure 2.13.
Section 3.3.5
⑤
⑧
Base
Base sense
By way of base resistance, it connects with the base (B) of the
socket of
①
or
②
in Figure 2.13.
It connects with the base (B) of the socket of
①
or
②
in Figure
2.13.
Section 3.3.5
⑥
⑨
Emitter
Emitter sense
It connects with the emitter (E) of the socket of
①
or
②
in Figure
2.13.
Section 3.3.5