Capter 3
Function Setting and Basic Operation
Curve Tracer CS-3000 Series Instruction Manual
3-93
3.11.4.1
Saving SETUP, TRACE, or REF data
[Procedure 1] Saving SETUP data into Internal Memory
(1) Press SAVE/RECALL key. SAVE/RECALL/DELETE menu appears.
(2) Press F1 key on the right of the submenu "SAVE."
(3) Press F1 key on the right of the submenu "SETUP/TRACE/REF" and
select "SETUP".
(4) Press F2 key on the right of the submenu "TO" and select the internal
memory icon .
(5) The submenu changes to "SETUP NO." and the number is enclosed by
the blue frame. SETUP number can be set.
(6) Rotate FUNCTION knob to set the number (1 to 256).
(7) Press F6 key on the right of the submenu "SAVE."
When being saved, "Save SETUP to #xx" (xx
=
SETUP NO) is displayed in the message area
on the upper part on the screen. Each time being saved, SETUP number increases one by
one.
[Procedure 2] Saving TRACE data into USB memory
(1) Press SAVE/RECALL key. SAVE/RECALL/DELETE menu appears.
(2) Press F1 key on the right of the submenu "SAVE."
(3) Press F1 key on the right of the submenu "SETUP/TRACE/REF" and
select "TRACE."
F3 submenu changes to "FILE NAME" and the 1st digit of the file name is
enclosed by the blue frame.
(4) Rotate FUNCTION knob to register the file name consisting of 8
characters; the first 4 alphabet characters (8-bit and
uppercase;!#$%&'()-@^_{}~) and 4 numeric 8-bit characters following
them.
To transfer between figures, press FUNCTION knob or F3 key.
(5) Press F4 key on the right of the submenu "DATA FORMAT" and select "BINARY" or "TEXT."
Text data cannot be recalled.
(6) Press F6 key on the right of the submenu "SAVE."
When being saved, "Save SETUP to ABCD0001.WFM" or "Save SETUP to ABCD0001.CSV"
is displayed in the message area of the upper part in the screen.
Each time being saved, FILE NAME number increases one by one.
If there is the file with the same file name when saving the file, the confirmation screen to
overwrite is displayed.
Ex.: CONFIRMATION
File "ABCD0001.WFM" already exists. Do you overwrite? Push [OK] button to overwrite. Push
[CANCEL] button to abort.
Press [OK] key to overwrite the file. Press [CANCEL] key to abort.