Chapter 4 Characteristics Measurement
4-22
Curve Tracer CS-3000 Series Instruction Manual
■
Measurement procedure
After step (19) in Section 4.1.2.1, I
C
vs. V
CE
characteristics is displayed.
(1) Rotate VERTICAL range counterclockwise to set to STEP.
(2) Rotate HORIZONTAL range clockwise to set to 2.00 V/div and COLLECTOR.
(3) Press STEP GENERATOR MENU key.
(4) Rotate FUNCTION knob to set NUMBER OF STEPS to 10.
The characteristics of the base current vs. the voltage between collector/emitter are displayed
with vertical 0.5div equal to one step.