Chapter 4 Characteristics Measurement
4-36
Curve Tracer CS-3000 Series Instruction Manual
■
Measurement procedure
(1) Press MODE/POLARITY key.
(2) Press F1 key and set to HIGH CURRENT.
(3) Press MAX PEAK POWER- WATTS key.
(4) Press F2 key and set to the measurement condition 4Kw (400A).
Measurement condition Ic=400 A in accordance with the performance table.
(5) Press MAX PEAK VOLTAS key.
(6) Press F3 key to check 40 V.
(7) Press STEP GENERATOR MENU key.
(8) Press F2 key to PULSE.
N-channel IGBT: plus bias
(9) Press F3 key to select VOLTAGE.
(10) Rotate FANCTION knob and set NUMBER OF STEPS to 0.
(11) Rotate STEP AMPLITUDE knob to set to 2V/div.
(12) Rotate OFFSET knob to set the measurement condition V
GE
.
Set the measurement condition V
GE
in accordance with the performance table.