Capter 2
Name and Function of Each Part (Outline)
2-22
Curve Tracer CS-3000 Series Instruction Manual
Caution!
-
Do not apply a voltage exceeding 1 kV or a current exceeding 1 A to each terminal of the
test adapter CS-506.
If 1 kV or 1 A is exceeded, test adapter or a measured device might be damaged.
Table 2.17 Name and function of each part on test adapter CS-506 (TO-252-5 type adapter)
No.
Name
Description
Ref.
①
Socket for element
It corresponds to the element of package TO-252-5 type.
It opens and shuts the lid of the socket, and mounts the element
corresponding to arbitrary pin assign.
Section
3.3.7
②
⑤
GATE
GATE SENSE
It connects with the pin set to GATE of the socket of
①
in
Figure 2.19.
Section
3.3.7
③
⑥
DRAIN
DRAIN SENSE
It connects with the DRAIN of the socket of
①
in Figure 2.19.
Section
3.3.7
④
⑦
SOURCE
SOURCE SENSE
It connects with the pin set to SOURCE of the socket of
①
in
Figure 2.19.
Section
3.3.7