Chapter 4 Characteristics Measurement
4-38
Curve Tracer CS-3000 Series Instruction Manual
After checking WAVE waveform
(25) Press F1 key and set VIEW MODE to TRACE.
(26) Press ACQUIRE key.
(27) Press F1 key and rotate FUNCTION knob to set SWEEP STEPS.
(28) Press F2 key and rotate FUNCTION knob to set PULSE INTERVAL.
(29) Rotate VARIABLE key again and apply Ic to the device as required.
(30) Press SWEEP key. (For SWEEP mode, see Section 3.8.)
(31) Sweep the range from VARIABLE setting value to 0
%
.
Saturation voltage V
CES
between collector/emitter of IGBT
Indicating collector being swept
Summary of Contents for CS-3000 Series