Capter 3
Function Setting and Basic Operation
3-100
Curve Tracer CS-3000 Series Instruction Manual
3.11.5 CONFIGURATION key
CONFIGURATION key sets connection between the collector/base/emitter terminals and the
COLLECTOR SUPPLY, STEP GENERATOR or Common.
Change to CONFIGURATION settings initializes VARIABLE setting values (see Section 3.7.1.2).
Pressing CONFIGURATION key
④
allows "CONFIGURATION" menu to appear.
Pressing function keys (F1 to F6) allows the internal connection state to change as shown below:
The comment on the connection status is displayed on the lower-right side of the internal
connection diagram.
Internal connection diagram
Destination of measured device C/B/E
F1
C: Collector Supply
B: Step Generator
E: Common
F2
C: Collector Supply
B: Open
E: Common