3.10.3 VIEW/PULSE key
Pressing VIEW/PULSE key
③
allows "VIEW MODE / PULSE" menu to appear.
If the collector supply is set to HIGH CURRENT MODE or if the step generator is set to PULSE,
the pulse width and the measurement point can be set.
STEP measurement
PULSE measurement
3.10.3.1
Submenus and functions
F1, F2, F3 settings
F1 setting allows F2 and F3 to change between Enable/Disable.
Submenu
Function description
F1 VIEW MODE TRACE
WAVE
Displays the trace waveform with
VERTICAL measurement value
used for the vertical axis and with
HORIZONTAL measurement
value used for the horizontal axis.
Displays the waveform with the measurement value
use for the vertical axis and with the time used for the
horizontal axis.
F2 WAVE 1st
Disabled
F2 key is used to select Ic
→
Vce
→
Vbe of the
waveform in sequence which is displayed on the
WAVE 1st (upper side). Rotating FUNCTION knob
allows waveform range to be set.
F3
WAVE 2nd
Disabled
F3 key is used to select Ic
→
Vce
→
Vbe of waveform
in sequence which is displayed on the WAVE 2nd
(lower side). Rotating FUNCTION knob allows
waveform range to be set.
* For F4 to F5, see
F4 and F5 settings
(page 3-44). F6 is not used.
Note 1: If HORIZONTAL source is set to STEP, Vce
=
500V/div and Vbe
=
5.00V/div are fixed.
Note 2: If VERTICAL source is set to STEP, Ic
=
2.00A/div is fixed.
F1
F2
F3
F4
F5
F6
F1
F2
F3
F4
F5
F6