Capter 3
Function Setting and Basic Operation
3-66
Curve Tracer CS-3000 Series Instruction Manual
3.11.1.4
SYSTEM SETUP submenu
Pressing F4 key allows the submenu "SYSTEM SETUP" to appear.
NO.
Submenus
Function description
F1
BEEP
When set to ON, the operation tone and error tone sound for
key/knob operation.
When set to OFF, no tone sound.
F2
POWER ON
SETUP
DEFAULT: This instrument starts with the initial state.
LAST:
This instrument starts with the state when it is powered
off previously.
F3
PANEL LOCK
The panel key can be invalidated excluding the
ENABLED/DISABLED breaker switch of collector supply output
(HIGH VOLTAGE/HIGH CURRENT) and the FUNCTION key to
PANEL LOCK and the CLEAR key.
By pushing the F3 key one by one, on/off of the panel lock function
is selected.
:O
N
state. The panel key becomes invalid excluding the
above-mentioned switch and the key.
:OFF state. All switches and the panel keys become
effective.
F4
HEADER SETUP The HEADER SETUP submenu opens by pushing the F4 key.
The selection of the type, the date, and the time for the header
displayed on the right of the screen can be set.
For how to set, see Section 3.11.1.4.1.
F5
MEASUREMENT
SETUP
The MEASUREMENTSETUP submenu opens by pushing the F5
key.
For how to set, see Section 3.11.1.4.2.
F6
(Not used)
F1
F2
F3
F4
F5
F6