Chapter 4 Characteristics Measurement
4-26
Curve Tracer CS-3000 Series Instruction Manual
(7) Press STEP GENERATOR MENU key.
(8) Press F1 key to select "-STEP."
(9) Press F3 key to select "VOLTAGE."
FET: voltage bias
(10) Rotate FUNCTION knob to set NUMBER OF STEPS to 6.
(11) Rotate STEP AMPLITDE knob to set to
−
200 mV/div.
Measurement condition (V
DS
=−
10 V, V
GS
=
0 V), V
GS
=
0 V at
−
200mV
×
5 step, in accordance
with the performance table.
Range of measured drain current I
DSS
at this time: 0.3 to 6.5mA
(12) Press OFFSET knob to set OFFSET to 0.
(13) Press MODE/POLARITY key.
(14) Press F1 key and set to HIGH VOLTAGE.
(15) Press F3 key and set RECTIFIED SINE to "
+
" (
+
full-wave rectified).
(16) Set VERTICAL to 500 μA/div and COLLECTOR.
Drain current I
DSS
=
−
0.3 to 6.5 mA in accordance with the performance table.
Make proper setting in accordance with device measurement value.
(17) Set HORIZONTAL to 2.00 V/div and COLLECTOR.
5 div. on screen equal to 2.00 V setting under measurement condition V
DS
=
10 V in
accordance with the performance table.