Capter 2
Name and Function of Each Part (Outline)
2-24
Curve Tracer CS-3000 Series Instruction Manual
Caution!
-
Do not apply a voltage exceeding 100 V or a current exceeding 0.5 A to each terminal of
the test adapter CS-507 and CS-509.
If 100 V or 0.5 A is exceeded, test adapter a measured device might be damaged.
Memo
- It is not possible to measure at the same time though two kinds of devices to be measured
can be installed in test adaptor CS-507 and CS-509. Measure by installing only one device
to be measured.
Table 2.18 Name and function of each part on test adapter CS-507 (SC-70-3 type adapter)
No.
Name
Description
Ref.
①
Socket for element
SDG
It corresponds to the element of package SC-70-3 (SOT-323-3)
type.
It mounts the array element of SOURCE, DRAIN and GATE
from the interior on the socket.
Section
3.3.8
②
Socket for element
GDS
It corresponds to the element of package SC-70-3 (SOT-323-3)
type.
It mounts the array element of GATE, DRAIN and SOURCE
from the interior on the socket.
Section
3.3.8
⑤
⑧
GATE
GATE SENSE
It connects with the pin set to GATE of the socket of
①
,
②
in
Figure 2.21.
Section
3.3.8
⑥
⑨
DRAIN
DRAIN SENSE
It connects with the pin set to DRAIN of the socket of
①
,
②
in
Figure 2.21.
Section
3.3.8
⑦
⑩
SOURCE
SOURCE SENSE
It connects with the pin set to SOURCE of the socket of
①
,
②
in Figure 2.21.
Section
3.3.8
Table 2.19 Name and function of each part on test adapter CS-509 (SC-59A type adapter)
No.
Name
Description
Ref.
③
Socket for element
SDG
It corresponds to the element of package SC-59A (SOT-23-3)
type.
It mounts the array element of SOURCE, DRAIN and GATE
from the interior on the socket.
Section
3.3.8
④
Socket for element
GDS
It corresponds to the element of package SC-59A (SOT-23-3)
type.
It mounts the array element of GATE, DRAIN and SOURCE
from the interior on the socket.
Section
3.3.8
⑤
⑧
GATE
GATE SENSE
It connects with the pin set to GATE of the socket of
③
,
④
in
Figure 2.22.
Section
3.3.8
⑥
⑨
DRAIN
DRAIN SENSE
It connects with the pin set to DRAIN of the socket of
③
,
④
in
Figure 2.22.
Section
3.3.8
⑦
⑩
SOURCE
SOURCE SENSE
It connects with the pin set to SOURCE of the socket of
③
,
④
in Figure 2.22.